HVAC-Talk: Heating, Air & Refrigeration Discussion banner
41 - 60 of 63 Posts
Are any digital refrigerant manifold w/ vacuum reading - trustworthy?
No. Unless it has a separate sensor like the YJ's.

The two biggest enemies to a micron gauge are positive pressure and oil. If you're using your manifold for anything other than an art piece - the internal micron sensor will be exposed to large amounts of oil and several hundred pounds of positive pressure per day.

I just can't see how any digital gauge with a built in micron gauge can stay clean (free of oil) and remain accurate.
 
I just got the Testo 557 and thought I would look to see the experiences of others.

Brent I can see your reasoning in thinking that the pressure and oil can have bad effects on the sensors. If this is the case,then I'm thinking that my new unused set should be more accurate. If not,then there must be other issues. Testo recommends blowing the gauge ports out with compressed air occasionally.

The only thing I have to compare is a yellow jacket digital I've had now for quite some time. I don't have the most faith in it either.

I have never pulled a vac seperate from my gauges. I can see the benefit though.

I think the thing that has me the most concerned is in reading the Testo manual it says that if the gauges are ever dropped or even as much as bumped hard, then you need to send them back for recalibration....really?.....maybe I'm a little rammy sometimes but even with a girl touch sometimes things get bumped in our trade. I can't see me not bumping this thing or NOT at one point seeing it come off the place I hooked it and drop. I'm not talking a 10ft. fall here, but my analog set sometimes come off and fall say 18" onto the dirt.

I'm almost afraid to touch this baby:D
 
If your old one is acting goofy try a new battery. I don't think the low battery indicator worked so well on those old ones and they drift around with low power.
 
Save
I think the thing that has me the most concerned is in reading the Testo manual it says that if the gauges are ever dropped or even as much as bumped hard, then you need to send them back for recalibration....really?.....maybe I'm a little rammy sometimes but even with a girl touch sometimes things get bumped in our trade. I can't see me not bumping this thing or NOT at one point seeing it come off the place I hooked it and drop. I'm not talking a 10ft. fall here, but my analog set sometimes come off and fall say 18" onto the dirt.

I'm almost afraid to touch this baby:D
Actually, in my Testo 570 manual it tells you that annual recalibrating is recommended. As far as dropping it, here's what it says in my manual:

If the measuring instrument falls or another comparable
mechanical load occurs, the pipe sections of the refrigerant
hoses may break. The valve positioners may also be damaged,
whereby further damage to the interior of the measuring
instrument may occur that cannot be identified from the outside.
The refrigerant hoses must therefore be replaced with new,
undamaged refrigerant hoses every time the measuring
instrument falls or following any other comparable mechanical
load. Send the measuring instrument to Testo Customer
Service for a technical check for your own safety.

Since the manifold is the heaviest part of the unit, it would make sense that it would hit the ground first. I still dont use my 570 for vacuum at all. I use the Appion CRT's, the large Appion hoses and my pump for vacuum.
 
I still wish I had never heard of Testo. :.02:
Actually, in my Testo 570 manual it tells you that annual recalibrating is recommended. As far as dropping it, here's what it says in my manual:

If the measuring instrument falls or another comparable
mechanical load occurs, the pipe sections of the refrigerant
hoses may break. The valve positioners may also be damaged,
whereby further damage to the interior of the measuring
instrument may occur that cannot be identified from the outside.
The refrigerant hoses must therefore be replaced with new,
undamaged refrigerant hoses every time the measuring
instrument falls or following any other comparable mechanical
load. Send the measuring instrument to Testo Customer
Service for a technical check for your own safety.

Since the manifold is the heaviest part of the unit, it would make sense that it would hit the ground first. I still dont use my 570 for vacuum at all. I use the Appion CRT's, the large Appion hoses and my pump for vacuum.
 
Every time i read these posts, i find myself saying "buy some digicools already". They aren't the sexiest, but at least you can drop them and not worry about needing a re-build. So what it you need a separate micron gauge, at least it won't be subjected to all the other stuff your gauges go through when you aren't pulling a vac.

To me sounds like testo owners are some of the best digicool sales people.

Sent from my GT-I9000M using Tapatalk 2
 
That or Sman.
Every time i read these posts, i find myself saying "buy some digicools already". They aren't the sexiest, but at least you can drop them and not worry about needing a re-build. So what it you need a separate micron gauge, at least it won't be subjected to all the other stuff your gauges go through when you aren't pulling a vac.

To me sounds like testo owners are some of the best digicool sales people.

Sent from my GT-I9000M using Tapatalk 2
 
I got the Testo 570's and love them. They work fantastic and are very accurate! I don't use them to pull a vacuum that what I bought my Bluvac for. Highly recommend the Testo 570's and the Bluvac!!


Sent from my iPhone using Tapatalk
 
41 - 60 of 63 Posts
You have insufficient privileges to reply here.